Scaling up a purification process is essential when large quantities of materials are required. Solid-liquid chromatographic processes often require method redevelopment as purification scale increases. However, for HPCCC, load mass and volume-related dimensions scale linearly in proportion to the ratio of the column volumes being scaled.
The data below demonstrate a 42-fold scale up of the four component mixture comprising 1: Methyl 2-acetamido- 5-bromobenzoate; 2: Warfarin; 3: 4-Bromobenzamide; 4: Dipyridamole.
| Instrument | Mini | Spectrum | Midi |
| Column | analytical | semi-preparative | preparative |
| Column Volume (mL) | 21 | 134 | 912 |
| Nominal Scale Up Factor | 1x | 6x | 42x |
| Flow Rate (mL/min) | 1 | 6 | 42 |
| Stationary Phase Retention (%) | 84 | 89 | 88 |
| Loading (g) | 0.2 | 1.2 | 8.4 |
| Loading Volume (mL) | 2 | 12 | 82 |
HEMWat SS 17 (Hexane:Ethyl acetate:Methanol:Water,1:1:1:1, v/v/v/v), NP mode elution. In all three experiments, following 30 minutes of ‘classical' elution, the column contents were extruded by pumping LP through the column